Evaluation of the Performance of Automated Haematology Analyser Dymind DH73 Compared to Sysmex XN 1000 System
Keywords:
Automated haematology analysers, correlation, Dymind DH73, haematological parameter, Sysmex XN-1000Abstract
Many types of automated haematology analysers are available for the used in clinical laboratories. It is reasonable to assume that a newly acquired piece of diagnostic equipment would run as intended, as manufacturers perform their own validation testing to prove intended use prior to launching a product in the market. Thorough validation testing on all new haematology analysers must be performed to ensure patient safety. This study was carried out to validate Dymind DH73, the automated haematology analysers in the Hospital Universiti Sains Malaysia setting. Blood samples send for complete blood count were selected randomly (n = 40) from healthy subjects and those who have different blood disorders. Blood specimens and quality control materials were analysed on the Dymind DH73 to evaluate precision, carry over and linearity. For correlation, we used the Sysmex XN-1000 as the comparative method. The study showed very good correlation (R > 0.9) between Dymind DH73 and Sysmex, XN-1000 in the parameters such as white blood cell, red blood cell, hemoglobin, platelet, nucleated rbc, neutrophil, lymphocyte, monocyte, eosinophil, and low correlation for basophil. In conclusion, the Dymind DH73 displayed a satisfactory performance with respect to precision, linearity and carry over. The performance of the Dymind DH73 analyser was good and compared favorably with the Sysmex XN-1000.
References
S.O. Ike, T. Nubila, E.O Ukaejiofo, I.N. Nubila, E.N. Shu, I. Ezema (2010, April). Comparison of haematological parameters determined by the Sysmex KX – 2IN automated haematology analyzer and the manual counts. BMC Clinical Pathology [Online]. 10(1). Available: https://pubmed.ncbi.nlm.nih.gov/20416068/
M. Velizarova, T. Yacheva, M. Genova, D. Svinorov (2021, September). Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system. J Med Biochem [Online] 40(4):367-377. Available: https://pubmed.ncbi.nlm.nih.gov/34616226/
D. Dave, A.N. Pandya (2014, March). Comparative Study of Four Hematology Analyzers. J Evol Med Dent Sci [Online] 3(12):3107-3113. Available: http://dx.doi.org/10.14260/jemds/2014/2251
R.K. Bhola, C. Fudaly, S. Rastogi (2024, April). A Comparative Evaluation of Performance of Sysmex XN 3000 and Horiba Yumizen H2500 Automated Complete Blood Count Analysers. Indian J Hematol Blood Transfus [Online] 40(2):303-314. Available: https://pubmed.ncbi.nl,.nih.gov/38708164/
P. Pusparini, A. Alvina (2022, September). Performance Comparison of Dymind DH-76 and Sysmex Xn-1000 Automated Hematology Analyzers. Indones J Clin Pathol Med Lab [Online] 257-262. Available: https://doi.org/10.24293/ijcpml.v28i3.1907.
M. Bruegel, D. Nagel, M. Funk, P Fuhrmann, J. Zander, D. Teupser (2015, January). Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-5000, and Sysmex XN-2000. Clin Chem Lab Med [Online] 53(7):1057-71. Available: https://doi.org/10.1515/cclm-2014-0945
E. Schapkaitz, S. Raburabu (2018, March). Performance evaluation of the new measurement channels on the automated Sysmex XN-9000 hematology analyzer. Clin Biochem [Online] 53:132-138. Available: https://doi.org/10.1016/j.clinbiochem.2018.01.014
C. Briggs, N. Culp, B. Davis, G. d’Onofrio, G. Zini, S.J. Machin (2014, March) ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting. Int J Hematol [Online] 36(6):613-27. Available: https://doi.org/10.1111/ijlh.12201
Products & Solutions, Dymind Biotechnology Co., Ltd. Shenzhen, China. [Online]. Available: https://www.dymind.com/en-US/products_and_solutions
S. Incir, K.E. Palaoglu. (2020, November). Evaluation of the performance of Sysmex XN-3100 automated hematology analyzer regarding the Sysmex XE-2100 and microscopic examination. Turk J Biochem. [Online]. pp. 29-37. Available: https://doi.org/10.1515/tjb-2020-0004
XN-1000 / 2000 XN-Series Automated Haematology Analysers Shaping Haematology, Sysmex Corporation, Kobe, Japan. [Online]. Available: https://www.sysmex-ap.com/wp-content/uploads/2020/08/XN-1000_2000-Brochure.pdf
S.S. Kaplan, K. Johnson, N. Wolfe, W. Brown, M. Keeny, L. Gray-Statchuk, et al. (2004, July) Performance characteristics of the Coulter LH 500 hematology analyzer. Lab Hematol. [Online] 10(2):76-87. Available: https://pubmed.ncbi.nlm.nih.gov/15224763/
Downloads
Additional Files
Published
Issue
Section
License
JBCS Publication Ethics
JBCS is committed to ensure the publication process follows specific academic ethics. Hence, Authors, Reviewers and Editors are required to conform to standards of ethical guidelines.
Authors
Authors should discuss objectively the significance of research work, technical detail and relevant references to enable others to replicate the experiments. JBCS do not accept fraudulent or inaccurate statements that may constitute towards unethical conduct.
Authors should ensure the originality of their works. In cases where the work and/or words of others have been used, appropriate acknowledgements should be made. JBCS do not accept plagiarism in all forms that constitute towards unethical publishing of an article.
This includes simultaneous submission of the same manuscript to more than one journal. Corresponding author is responsible for the full consensus of all co-authors in approving the final version of the paper and its submission for publication.
Reviewers
Reviewers of JBCS treat manuscripts received for review as confidential documents. Therefore, Reviewers must ensure the confidentiality and should not use privileged information and/or ideas obtained through peer review for personal advantage.
Reviews should be conducted based on academic merit and observations should be formulated clearly with supporting arguments. In cases where selected Reviewer feels unqualified to review a manuscript, Reviewer should notify the editor and excuse himself from the review process in TWO (2) weeks time from the review offer is made.
In any reasonable circumstances, Reviewers should not consider to evaluate manuscripts if they have conflicts of interest (i.e: competitive, collaborative and/or other connections with any of the authors, companies, or institutions affiliated to the papers).
Editors 
Editors should evaluate manuscripts exclusively based on their academic merit. JBCS strictly do not allow editors to use unpublished information of authors  without the written consent of the author. Editors are required to take appropriate responsive actions if ethical complaints have been presented concerning a submitted manuscript or published paper.
CONFLICT OF INTEREST
Journal of Biomedical and Clinical Sciences requires authors to declare all competing interests in relation to their work. All submitted manuscripts must include a ‘competing interests section at the end of the manuscript listing all competing interests (financial and non-financial). Where authors have no competing interests, the statement should read ,The authors have declared that no competing interests exist. Editors may ask for further information relating to competing interests.
Editors and reviewers are also required to declare any competing interests and will be excluded from the peer review process if a competing interest exists. Competing interests may be financial or non-financial. A competing interest exists when the authors interpretation of data or presentation of information may be influenced by their personal or financial relationship with other people or organizations. Authors should disclose any financial competing interests but also any non-financial competing interests that may cause them embarrassment if they were to become public after the publication of the article.
HUMAN AND ANIMAL RIGHTS
All research must have been carried out within an appropriate ethical framework. If there is suspicion that work has not taken place within an appropriate ethical framework, Editors will follow the Misconduct policy and may reject the manuscript, and/or contact the author(s) institution or ethics committee. On rare occasions, if the Editor has serious concerns about the ethics of a study, the manuscript may be rejected on ethical grounds, even if approval from an ethics committee has been obtained.
Research involving human subjects, human material, or human data, must have been performed in accordance with the Declaration of Helsinki and must have been approved by an appropriate ethics committee. A statement detailing this, including the name of the ethics committee and the reference number where appropriate, must appear in all manuscripts reporting such research. Further information and documentation to support this should be made available to Editors on request.
Experimental research on vertebrates or any regulated invertebrates must comply with institutional, national, or international guidelines, and where available should have been approved by an appropriate ethics committee. The Basel Declaration outlines fundamental principles to adhere to when conducting research in animals and the International Council for Laboratory Animal Science (ICLAS) has also published ethical guidelines.
A statement detailing compliance with relevant guidelines (e.g. the revised Animals (Scientific Procedures) Act 1986 in the UK and Directive 2010/63/EU in Europe) and/or ethical approval (including the name of the ethics committee and the reference number where appropriate) must be included in the manuscript. The Editor will take account of animal welfare issues and reserves the right to reject a manuscript, especially if the research involves protocols that are inconsistent with commonly accepted norms of animal research. In rare cases, Editors may contact the ethics committee for further information.
INFORMED CONSENT 
For all research involving human subjects, informed consent to participate in the study should be obtained from participants (or their parent or guardian in the case of children under 16) and a statement to this effect should appear in the manuscript, this includes to all manuscripts that include details, images, or videos relating to individual participants.
DATA SHARING POLICY
JBCS strongly encourages that all datasets on which the conclusions of the paper rely should be available to readers. We encourage authors to ensure that their datasets are either deposited in publicly available repositories (where available and appropriate) or presented in the main manuscript or additional supporting files, in machine-readable format (such as spreadsheets rather than PDFs) whenever possible
Authors who do not wish to share their data must state that data will not be shared, and give the reason.
COPYRIGHT NOTICE
The JBCS retains the copyright of published manuscripts under the terms of the Copyright Transfer Agreement. However, the journal permits unrestricted use, distribution, and reproduction in any medium, provided permission to reuse, distribute and reproduce is obtained from the Journal's Editor and the original work is properly cited.
While the advice and information in this journal are believed to be true and accurate on the date of its going to press, neither the authors, the editors, nor the publisher can accept any legal responsibility for any errors or omissions that may be made. The publisher makes no warranty, express or implied, with respect to the material contained herein.
Copyright (c) 2023 Journal of Biomedical and Clinical Sciences (JBCS)
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.